Cleaning and Sterilization Instructions for
Sterrad® 100S / 200 / NX / 100NX

Suggested cleaning procedure for the following Koven Technology surgical probe(s):

The probe models listed below are compatible with Sterrad® 100S / 200 / NX / 100NX
VRP Series
PP8M05N5A
P5M05S8A
VDP-8A / VDP-10A
PP10M05N5A
P8M05S8A
LRP-8
P5M05N5A
P10M05S8A
FDP Series
P8M05N5A

VGP-8
P10M05N5A

The probe models listed below are compatible with Sterrad® 100S / 200 / NX only
HP10M05N5A / HP20M05N5A HP10M1N5AF / HP20M1N5AF  
CRP Series
NRP Series

STERRAD STERILIZATION SYSTEM GUIDELINES:
The probes listed above may be sterilized with the Sterrad 100S, Sterrad 200, Sterrad NX and/or Sterrad 100NX (where indicated) sterilization systems (see charts above).  The general processing instructions in the “Sterrad 100S, Sterrad 200, Sterrad NX and/or Sterrad 100NX Sterilization System Operator’s Manual(s)” should be followed. Probes must be cleaned to remove contaminating substances prior to sterilization. After each use, perform initial cleaning within the sterile field by wiping the probe free of blood and body fluids.

CLEANING PROCESS:  
Wipe probe & cable down with a damp cloth using an enzymatic cleaner. Wipe the probe and cable again using a damp cloth of warm water or alcohol before placing in the sterilizer. 

STERILIZATION:   Per recommended Sterrad 100S, Sterrad 200, Sterrad NX and/or Sterrad 100NX procedures using a Sterrad brand or Sterrad compatible trays.

WARNING:  Sterilize according to instructions for the Sterrad 100S, Sterrad 200, Sterrad NX, and/or Sterrad 100NX Model(s).  Do not sterilize the Doppler probes listed above by steam autoclaving. Steam sterilization is likely to damage the plastic probe tip and piezo-electric crystals.  Utilize Koven/Hadeco probe number ACP-8 for steam autoclaving. Do not soak or submerse the probe in liquidDo not use Cidex.

Failure to use approved cleaning & sterilization procedures may void the warranty

Sterrad is a registered trademark of Advanced Sterilization Products, a Johnson & Johnson Company